Numerical Design of Dual Resonant Microstrip Surface RF Coil for MRI Application
Penulis:Â Mahendra, Adhi;Â Basari;Â Rahardjo, Eko Tjipto
Informasi
JurnalProceedings - CAMA 2019: IEEE International Conference on Antenna Measurements and Applications, 2019 IEEE Conference on Antenna Measurements & Applications (CAMA)
PenerbitInstitute of Electrical and Electronics Engineers Inc., IEEE
Halaman320 - 322
Tahun Publikasi2019
ISBN978-172812396-7
Jenis SumberScopus
Sitasi
Scopus: 2
Google Scholar: 2
PubMed: 2
Abstrak
Magnetic resonance imaging (MRI) has three main components, namely the main magnet, gradient coil, and RF coil. RF coils play an important role as recipients of RF signals from the emission of magnetic resonance and RF excitation into atomic nuclei of the human body. In this paper, we propose a design of dual resonant microstrip RF coil that operates at 1.5 T and 3 T MRI. The proposed coil is simple structure for surface RF coil and capable of working at frequencies of 63.8 MHz and 127.6 MHz. The simulation results show that the reflection coefficient (S11) is less than -10 dB either without or with the human phantom model at both operating frequencies. As for magnetic field distribution, the field is more homogeneous at the lower operating frequency compared to the higher one. The computed peak specific absorption rate (SAR) is obtained by about 0.56 W/kg and 0.91 W/kg at 63.8 MHz and 127.6 MHz, respectively. © 2019 IEEE.
Dokumen & Tautan