Modified calibration protocols in electron beam dosimetry: comparison with IAEA TRS-398 and AAPM TG-51

Penulis: Cica Yulinar, Muhamad Iqbal Assegab, Wahyu Edy Wibowo, Supriyanto Ardjo Pawiro
Informasi
JurnalBiomedical Physics and Engineering Express, Biomedical Physics & Engineering Express
PenerbitBiomedical Physics & Engineering Express 9 (5), 055008, 2023, Institute of Physics, IOP Publishing
Volume & EdisiEdisi 5
Tahun Publikasi2023
ISSN20571976
Jenis SumberGoogle Scholar
Sitasi
Scopus: 5
Google Scholar: 5
PubMed: 5
Abstrak
This study aimed to compare absolute calibration outputs based on the protocols of the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)-398, the American Association of Physicists in Medicine (AAPM) Task Group (TG)-51, and modified calibration approach. The electron beam output calibration followed the IAEA TRS-398 and AAPM TG-51 protocols, both of which required cylindrical chambers and parallel plates. However, the use of cylindrical chambers is not recommended at low energies because of the large fluence-correction factor. TG-51 recommended cross-calibration of the parallel-plate chamber against the cylindrical chamber in a high-energy electron beam. In 2020, an electron beam dosimetry modification was introduced that used a cylindrical ionisation chamber at all energies and updated the data for beam quality conversion factors. This modification provided a lower …
Dokumen & Tautan